| Business Type | Manufacturer, Supplier |
| Country of Origin | India |
| Warranty | 1 Year |
| Application | Industrial Use |
| Click to view more | |
Product Details
The sample solution is aspirated through an auto-miser. Sample, Air and the fuel are mixed in the mixing chamber which is then sprayed as a very fine mist into the flame. The color of the flame is changed depending upon the concentration of elements present. Radiations from the flame pass through the sensing system and specific narrow band interference filter which permits only the characteristic radiation to pass to the photo-detector. The output of the photo-detector is then processed by the microcontroller and the final results are displayed on graphical LCD display.
In both models 5-Point calibration facility and non-linearity correction is provided through automatic curve fitting software. Facility has been provided for re-standardization with a single standard, thereby, eliminating the need of the multiple standards once a calibration has been performed.
Salient Features:
Specification
| Mode | General | Serum | Urine |
| Range | Na: 0 - 100 ppm Upto 250 meq/l, 1:100 dilK: 0 - 100 ppm Upto 250 meq/l, 1:100 dilCa: 0 - 100 ppm Upto 250 meq/l, 1:100 dilLi: 0 - 100 ppm Upto 250 meq/l, 1:100 dil | Na: 0 - 200 1:100 dilK: 0 - 10 1:100 dil-Li: 0 - 2 1:5 dil | Na: 0 - 250 1:100 dilK: 0 - 100 1:100 dilCa: 0 - 10 1:2 dil- |
| Sensitivity | General: Na: 0.5 ppm, K: 0.5 ppm, Li: 0.5 ppm, Ca: 15 ppm | ||
| Filters | Na & K supplied (Ca & Li are optional at extra cost) | ||
| Resolution | 0.1 ppm/meq | ||
| Reproducibility | < 1% coefficient of variation for 20 consecutive samples using 10 ppm Na set as maximum standard | ||
| Curve Fit Accuracy | < 2% error when 3 ppm Na/K and 5 ppm Li are set as maximum standards | ||
| Display | 128 x 64 Graphical LCD Display | ||
| Average Time | In-built in software | ||
| Flame System | LPG and dry oil free air | ||
| Detector | Silicon photodiode | ||
| Calibration | Upto 5-Point calibration with curve fitting software | ||
| Gas Control | Adjustable with knobs | ||
| Flame Failure | Auto detection | ||
| Gas Cutoff | Automatic | ||
| Ignition System | Auto ignition | ||
| Atomizer | Axial flow type | ||
| Printer | Provision for attachment of dot matrix printer or HP laserjet/ inkjet (non-windows) printers with centronics parallel interface | ||
| RS-232C | Available (PC software at extra cost) | ||
| Power | 230 V + 10% AC, 50 Hz | ||
| Dimensions | 285 x 255 x 210 mm (L x B x H) (Approx.) | ||
| Weight | 7.5 Kg (Approx.) | ||
COMPRESSOR UNIT |
|||
| Air Supply | By oil free mini compressor unit with pressure regulator | ||
| Combustion Gas | LPG controlled by precision regulator | ||
| Power | 230 V + 10% AC, 50 Hz | ||
| Dimensions | 290 x 255 x 210 mm (L x B x H) (Approx.) | ||
| Weight | 8 Kg (Approx.) | ||
Hi! Simply click below and type your query.