EnglishFrenchGermanItalian

Microprocessor Conductivity Meter


Microprocessor Conductivity Meter

With the outstanding storage facility, we are capable to meet urgent and massive requirements of the customers. Also, we keep it updated to meet industry standards. Maintaining high standards of our extensive array of products, we focus on the quality of raw inputs and material used at the initial stages. With this, the final product comes out to be the finest.

 

Features :

  • Accurate result
  • Robust structure
  • Corrosion resistant
  • Easy to operate


Other details :

  • Highly sensitive & Accurate
  • Measurement Conductivity, TDS and Temperature
  • Storage up to 1000 samples
  • Printer attachment facility.


Technical Specifications

Function Cond. & Temp., TDS & Temp., SAL & Temp.Simultaneously
Display 16 character x 2 lines alphanumeric BL LCD
Power supply 230 V ± 10% 50 Hz AC
Conductivity -TDS-SAL  
Range Conductivity 0.00 uS/cm to 1000 mS Auto
Ranging TDS 0.0 ppm to 1000 ppt Auto
Ranging SAL 0.0 ppt to 50 ppt Auto
Resolution Conductivity 0.01 uS on the lowest range
TDS 0.1 ppm
SAL 0.1 ppt
Accuracy 0.5% of range 1 digit
TDS Factor 0.50(By default) & adjustable manually
SAL Factor 0.75(By default) & adjustable manually
Cell const. adjustment Digital 0.05 to 2.000 By default 1.000
Temp.compensation 0 to 70oC Auto with temp. probe/Manual with keys
Temp. Coefficient adj. 0.00% to 4.00% adjustable. By default 2.00%
Calibration Manual
Reference Temp 20 Degree C & 25 Degree C
Measuring Frequency 100 Hz & 1 KHz Auto Selection
Mode of Calibration By Value & By Measurement
Temperature
Range -200oC to +200oC
Resolution 0.1oC
Accuracy + 0.1% of range +1 digit
Calibration two point manual
Sensors PT100 probe
Data entry By six soft touch keys with audible sound
Storage 1000 samples
Real time display 24 Hour mode with Date
Interface Centronics & computer
Print format All Data Date Wise
Digital output Serial RS-232 Parallel Centronics
Accessories Conductivity Cell, Instruction Manual, Dust cover, Temp Probe

 

Appearance & specification subjected to change due to continuous development.


Print Format

Date Time Temp(DegC) Parameter Mode SP. No
39794 0.47 T= 020.0 C=00.00us/cm Manul 1
39794 0.47 T= 020.0 TDS=00.00ppm Manul 2
39794 0.47 T= 020.0 SAL=000.5ppt Manul 3